NEONATOLOGY ON THE WEB


Title Computerised Determination of Spontaneous Inspiratory and Expiratory Times in Premature Neonates During Intermittent Positive Pressure Ventilation. I: A New Technique
Author(s) J. N. Mockridge, J. S. Ahluwalia, and C. J. Morley
Source Arch-Dis-Child-Fetal-Neonatal-Ed, Vol. 71, No. 3, Pages 156-160
Publication Date Nov-94
Abstract A pneumatic capsule applied to the sub-xiphisternum provides a signal of spontaneous respiratory effort. Purpose-written software analyses this waveform to determine spontaneous inspiratory and expiratory times. Algorithms reject artefacts. The computerised system allows continuous measurement of spontaneous inspiratory and expiratory times to be made during intermittent positive pressure ventilation. The accuracy of the system was measured in 17 infants ventilated for respiratory distress syndrome: median (range) gestation and birth weight were 31 (25-38) weeks and 1195 (506-3260) g. The new technique can measure inspiratory time of individual breaths with a mean error of 5.4% and expiratory time with a mean error of 1.6%.


Return to Citation Index Page

This page was last modified on 12-10-1995.
Citation database was last rebuilt on 08-17-2022.
Neonatology on the Web / webmaster@neonatology.net