Chapple AJDC 1938
Controlling the External Environment of Premature Infants in an Incubator.
Dr. Charles C. Chapple (by invitation), Philadelphia.
American Journal of Diseases of Children, Vol. 56, pages 459-460, 1938
In an attempt to solve the problem of cross infection at the Children’s Hospital of Philadelphia, an incubator was built to circulate outdoor air. Ward air is excluded from the chamber containing the infant by (1) handling the baby through balloon cloth sleeves through which only the hands are inserted; (2) admitting bottles and diapers through an air lock which prevents the entrance of any current of ward air, and (3) maintaining a small positive pressure inside the chamber so that in any leakage which may take place air will pass from the chamber and not into it. In this incubator any desired temperature and humidity can be held with slight fluctuation, and special mixtures of gases, such as those with high oxygen concentrations, can be maintained.
An incubator was installed in a fifteen-bed ward of sick older infants, where cross infections were constantly occurring. The incubator attendants wore no masks or gowns but were careful to wash the hands before inserting them through the sleeves.
Under these conditions bacterial counts were taken by the Wells method of centrifuging the air against agar tubes. The ward air counts varied from 400 to 1,100 bacterial colonies per tube (representing 5 cubic feet [0.14 cubic meter] of air); the outdoor air, from 20 to 40 colonies, and the incubator air, from 2 to 20 colonies. The incubator was unoccupied when the tests were made, but it had been in the ward for one year and had been cleaned two weeks previously, when the last occupant was removed. The mechanism had been functioning for two hours, and it was maintaining a temperature of 90 F. and a relative humidity of 70 per cent.
Similar tests were made on an occupied incubator in the nursery of a maternity hospital, where the air in the ward averaged 90 colonies per tube and the outdoor air 23 colonies. The incubator in this ward averaged 9 colonies per tube.
While the total number of infants cared for in the incubator is not yet great enough (10 in the Children’s Hospital and 16 in maternity departments of other hospitals) to be conclusive evidence in favor of this method of caring for the premature baby, the evidence so far has been favorable. No infection of the respiratory tract has occurred in any infant in an incubator, and there has been only 1 case of transitory diarrhea, while in a control group of similar size under strict isolation at the Children’s Hospital there were 4 cases of fatal pneumonia and 1 of fatal diarrhea. The only death caused by infection in an incubator resulted from septicemia originating in an abscess at the site of repeated hypodermoclyses.
Cyanotic attacks were reduced in number by the use of high humidity and high oxygen concentration.
The diurnal fluctuation of temperature as recorded rectally was seldom as great as 1 degree F. Since the temperature is held so exactly to the desired level, the babies are left uncovered and unclothed. Thus they are free to exercise, which they do to a remarkable degree, and their respiration is unhampered by the weight of blankets.
Continuous observation of the baby in the nude state is permitted.
Last Updated on 11/10/26